Rosemount™ 65 RTD Temperature Sensor
As an industry-standard Pt-100 resistant temperature detector (RTD), the Rosemount 65 RTD Temperature Sensor is a single or dual element device designed per IEC 751 and capable of covering a wide range of temperatures.
This sensor is vibration resistant and the connection heads feature fast mounting and replacement. This RTD includes a calibration option of Class A or B accuracy or a calibration schedule option with Callendar-Van Dusen constants.
Rosemount™ 65 RTD Temperature Sensor
Specs
Time Constant / Response
9 seconds maximum required to reach 50% sensor response when tested in flowing water according to IEC 751:1983, Amendments 1 and 2
Insulation Resistance
1,000 MΩ minimum insulation resistance when measured at 500 Vdc and at room temperature
Calibration Options
IEC 751 Class A accuracy; Calibration schedule with Callendar-Van Dusen constants; MID custody transfer calibration
Resources
-
Certificates & Approvals
-
Data Sheets & Bulletins
-
Data Sheets & Bulletins
-
Drawings & Schematics
-
Drawings & Schematics
-
Drawings & Schematics
-
Drawings & Schematics
-
Drawings & Schematics
-
Manuals & Guides
Features
- Industry-standard sensor per IEC 751 with single or dual element design offers reliable and flexible performance
- Resilient sensor withstands high vibration industrial applications
- DIN-style sensor connection heads allow for quick mounting and replacement
- Sensor selections cover a wide range of temperatures from –321 to 1112°F (–196 to 600°C) to meet process demands
- Terminal block, flying leads and spring-loaded threaded adapters enable remote or integral mounting configuration
- Callendar-Van Dusen constants allow for transmitter-sensor matching and provide the highest level of accuracy
- Optional Class A accuracy ensures critical temperature measurement points meet tolerance standards and accuracy
- Option for thermowell assembly increases protection from corrosive environments
- Measurement Instrument Directive (MID) calibration options for custody transfer applications
- Global hazardous location approvals meet local requirements